To design a Standard Operating Procedure that describes the procedure for operation, and maintenance of In Vivo Imaging System Perkin.
2.0 SCOPE
This Standard Operating Procedure (SOP) is applicable for the operation, maintenance, and of In Vivo Imaging System Perkin for Pharmacology studies conducted at the Animal House Facility
3.0 RESPONSIBILITY
3.1 All personnel in the Pharmacology Department shall be responsible for implementation and compliance with the SOP.
4.0 DISTRIBUTION:
4.1 The Quality Assurance department is responsible for keeping SOP ‘Master Copy’ approved through the Quality Manager.
5.0 DEFINITION(S)
5.1 Definition(s)
5.1.1 Imaging of fluorescence in vivo has become the most commonly used radiological tool in all pre-clinical work however fluorescence sampling of cellular function is widely used in all aspects of biology, allowing the visualization of cellular and sub-cellular biological processes with spatial resolutions in the range from nanometers up to centimeters.
6.0 ABBREVIATION(S)
6.1.1 SOP: Standard Operating Procedure
6.1.2 LAF : Laboratory Animal Facility
6.1.3 PPEs : Personal Protective Equipment
6.1.4 UV : Ultra Violet
6.1.5 mm : Millimeter
6.1.6 e.g. : For Example
6.1.7 CCD : Charged-Coupled Device
6.1.8 Nm : Nanometer
6.1.9 CLS : Caliper Life Science
7.0 PROCEDURE
7.1 Principle
7.2 Operations
7.2.1 General start-up procedure
7.2.1.1 Sign on the log sheet once you enter the small Animal In Vivo Imaging Room.
7.2.1.2 Clean the lab benches, sample mat, and imaging chamber with 70% ethanol. Switch on the air purifier.
7.2.1.3 Switch on the monitor of the computer connected with IVIS Spectrum. (The computer and the software (Living Image) shall always be on, shutdown/restart of the system or software shall not be recommended)
Figure 1: Living Image
7.2.1.4 If the software is incidentally closed, launch the living image log in as CLS (Caliper Life Science), and then initialize the system. Wait until the system “Idle”.
Figure 2: CLS Log-In Screen
7.2.1.5 The camera's temperature shall be checked by clicking the green/red status box. The system shall be ready to use when the status box turns green, i.e. the camera temperature ≤ -87°C. Otherwise, wait unit system is ready.
Figure 3: Camera Temperature Screen
7.2.1.6 Prepare the sample while waiting.
7.2.1.7 Select Acquisition>Auto Save To select the destination for saving image data as user folder in D: Drive> User>
Figure 4: Acquisition Screen
7.2.1.8 In Field of View in the control panel, the user shall choose a calibrated position/area of the sample stage which are following:7.2.1.8.1 4 x 4 cm; (Closest to the camera; highest sensitivities)
7.2.1.8.2 6.5 x 6.5 cm;
7.2.1.8.3 1 3 x 13 cm;
7.2.1.8.4 22.5 x 22.5 cm; (Farthest from the camera; lowest sensitivities)
7.2.1.9 Open the imaging chamber door when the specimen is ready.
7.2.1.10 Place the Specimens or anesthetized animal on the sample mat (use a perforated plate whenever 3D Trans-Illumination Fluorescent acquisition is needed; otherwise use the 30cm x 30cm Sample Mat) and within the green laser grid on the stage. The cross indicates the center of the imaging area.
7.2.1.11 Close the imaging chamber door.
7.2.2 Basic operation for 2d bioluminescence imaging
7.2.2.1 In the control panel, check the box of the luminescent image and set up the acquisition parameters automatically or manually.
7.2.2.1.1 For automatic set-up, select auto in the exposure box, the software shall determine exposure time, binning, and F/stop settings,
7.2.2.1.2 To manually set up, adjust exposure time (0.5 sec to 5 min) according to the signal intensity
7.2.2.2 Then binning shall be adjusted if necessary as:
7.2.2.2.1 Small (4) for High resolution; Low sensitivity;
7.2.2.2.2 Medium (8);
7.2.2.2.3 Large (16) for Low resolution; High sensitivity;
7.2.2.3 Adjust F/stop if necessary. 1 for large aperture; 8 for small aperture
7.2.2.4 Adjust Focus if necessary. Input the subject height (in cm) of the specimen.
Figure 5: Acquisition Parameter for 2D Bioluminescence
7.2.2.5 Check the box of the photograph for a bright field image.
7.2.2.6 Acquire to capture the photography and luminescent image.
7.2.2.7 During the acquisition, fill in the sample information and click OK.
Figure 6: Sample Information
screen
7.2.2.8 Save the data in D: Drive>User>Your Folder if AutoSave is not activated.
7.2.2.9 Perform image display adjustment, ROI measurement, etc. analysis in the offline software (Please reserve spectrum offline session in FCF online booking system; free of charge) according to the living images analysis protocol.
7.3 Maintenance
7.3.1 General Cleaning and Finishing Procedures
7.3.1.1 Close all the images but do not exit the living image software.
7.3.1.2 Transfer image/data files using the FCF network server according to the data transfer instructions. All the browsers but not living images shall be closed after the transfer.
7.3.1.3 All user samples shall be removed from the imaging chamber. The stage and sample mat shall be cleaned with 70% ethanol and Kimwipes.
7.3.1.4 The bench shall be cleaned with 70% ethanol and a paper towel.
7.3.1.5 The system runs calibration at midnight for accurate intensity measurement for the next day's use. Please do
7.3.1.5.1 Not exit living image software but close all other software and images
7.3.1.5.2 Make sure the chamber door shall be closed properly.
7.3.1.5.3 Turn off the monitor only (NEVER turn off the computer or any parts of the system!!!)
7.3.1.6 Sign out in the log book according to the actual leaving time.
7.3.1.7 Turn off the light and close/lock the door when leaving.
8.0 PRECAUTION
8.1 All PPEs like aprons, gloves, masks, head caps, and shoe covers should be worn before using the instrument.
8.2 Press Keys softly.
8.3 Clean the equipment with a dry piece of robe or a lightly moistened one, do not use organic solutions.
8.4 Never use abrasive products or dissolvent.
8.5 Never pour water or liquids on the equipment.
8.6 Once you have finished using the equipment turn it off by using the main switch, clean, and check the equipment so that it is in optimal condition when you use it again.
8.7 Do not position the unit so that it is difficult to operate the main disconnect device.
8.8 Electrical outlets in the cabinet are restricted to 5 amps maximum current.
8.9 Do not use any detachable power cord that is not adequately rated for the unit.
8.10 Follow the rules of the animal house.
9.0 REFERENCE & ATTACHMENTS
9.1 Reference
9.1.1 Manufacturer manual in vivo imaging system perkin.
9.2 Format
Format – I Format for Log Book Maintenance
Format –II Format for Document History Sheet
10.0 REVISION HISTORY
END OF THE DOCUMENT
You may like to read these links:
1. List of All SOPs and Documents for the Microbiology Laboratory
2. List of All SOPs and Documents for In-vitro Laboratory
3. List of All SOPs and Documents for the In-vivo Laboratory.
4. List of All SOPs and Documents for Clinical Research
5. List of Instruments and Equipment Used In Laboratory Animal Facility
6. Animal Facility Design - Small Laboratory Animals (Rat, Mice, Rabbit, Guinea Pig)
7. List of Chemicals Needed for In-vivo Laboratory
8. List of Physiological Data for Small Laboratory Animals