STANDARD OPERATING PROCEDURE FOR CLEANING, OPERATION & CALIBRATION OF AUTOCLAVE | SOP FOR AUTOCLVAE

SOP FOR AUTOCLAVE

1.0              OBJECTIVE

             To lay down the procedure for Cleaning, Operation & Calibration of the Vertical Autoclave.

2.0              SCOPE

This SOP shall be applicable for the Cleaning, Operation & Calibration of the Vertical Autoclave in the Microbiology Laboratory.

3.0              RESPONSIBILITY

3.1              Microbiologists and above shall be responsible for the preparation of this SOP.

3.2              Manager-QA shall be responsible for the approval of this SOP.

4.0              ACCOUNTABILITY

Sr. Microbiologist

5.0              PROCEDURE

5.1              Cleaning

5.1.1        Ensure that the power supply to the vertical autoclave is switched ‘OFF’ and the plug shall be removed from the socket before cleaning.

5.1.2        Dust shall be removed from the surface of the vertical autoclave with a clean dry cotton cloth.

5.1.3        Water shall be drained out from the vertical autoclave.

5.1.4        Inside of the vertical autoclave shall be cleaned with a clean cotton cloth wetted with 70% I.P.A 

5.2              Operation

5.2.1        Ensure the water level in a vertical autoclave. If the level is below the minimum level, distilled water shall be added, and ensure that heating rods shall be dipped into the water.

5.2.2    Material to be sterilized shall be wrapped with brown paper/silver foil and stick a vertical autoclave indicator strip on the material and place the material inside the chamber of the vertical autoclave and inside wiring of Stainless steel.

5.2.3        Ensure that the lid of the vertical autoclave shall be closed and tighten the screws of the lid.

5.2.4      Ensure that the power supply of the autoclave is switched “ON” and allow the temperature to reach 100° C. Keep the steam vent open for 5 minutes for escaping the inside air. Close the steam vent and allow the temperature and pressure to attain 121°C & 15 psi respectively above atmospheric pressure for 15 minutes. Note down the time when temperature and/or pressure are achieved

5.2.5        After a holding time switch “OFF” the power supply of the autoclave.

5.2.6        Ensure when the pressure comes to zero, slowly open the vent to release the steam.

5.2.7        Open the screws of the lid and allow it to cool.

5.2.8        Ensure the color change in the indicator strip and take out the sterilized material from the inner chamber of the vertical autoclave.

5.2.9         Enter the cycle period in the daily usage logbook as per annexure I& II

5.3             Cycle Numbering

5.3.1        Allocate an Identification numbering for each cycle of that equipment separately.

5.3.2        A typical cycle number shall be as Q__/AC/MM/YY/000.

5.3.3        In the identification number Q-- shall be abbreviate for Quality Control, AC shall stand for autoclave cycle number, MM &YY shall stand for the month and last two digits of that year respectively, and 000 present three numerical serial number which starts from 001.

5.4              Internal Calibration Procedure

5.4.1        Self-adhesive autoclavable tape initially off-white in color, after sterilization changes to black or dark brown shall be affixed on a flask during sterilization by autoclave.

5.4.2        This shall be done for each & every cycle of sterilization.

5.4.3        Calibration of the Instrument:

·      External calibration shall be done for Instrument once in 12 months.

5.5              Precautions

5.5.1        Thermal gloves shall be used to open the autoclave.

5.5.2        Do not touch the vertical autoclave when it is hot.

6.0              REFERENCE

Not applicable

7.0              ANNEXURE

Annexure No.

Title of the Annexure

Format No.

  Annexure I

 Sterilization record of Autoclave

 

  Annexure-II

 Discarding record of Autoclave

 

8.0              ABBREVIATIONS :

SOP                                         :           Standard Operating Procedure

QCD                                       :           Quality Control Department

EQP                                        :           Equipment

QA                                          :           Quality Assurance

PSI                                          :           Pounds Per Square Inch

No.                                          :           Number

I.P.A                                       :           Isopropyl alcohol


                                                                   END OF DOCUMENTS

   ANNEXURE-II: DISCARDING RECORD OF AUTOCLAVE

Sr. No.

Date

Name of Item

No. of Cycle

Sterilization Cycle

Holding Time

Status of   Indicator

Done by

Autoclave started at

1210C&15psi

attain time

Completed at

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

You may like to read these links: 

1. List of All SOPs and Documents for the Microbiology Laboratory

2. List of All SOPs and Documents for In-vitro Laboratory

3. List of All SOPs and Documents for the Animal House Facility



                       

SHARE

Owner

Hi. I’m Writer of Researchsop.com. ’ ’ Please share these SOPs to all concern pharma people for their development. I like to fullfill the need of curious people. These things inspire me to make things looks better.

  • Image
  • Image
  • Image
  • Image
  • Image
    Blogger Comment
    Facebook Comment

0 comments:

Post a Comment