SOP FOR LAMINAR AIR FLOW | STANDARD OPERATING PROCEDURE FOR CLEANING, OPERATION & CALIBRATION OF LAMINAR AIR FLOW

  

1.0              OBJECTIVE

To lay down the procedure for cleaning, operation & calibration of laminar airflow.

2.0              SCOPE

This SOP shall be applicable for cleaning, operation & calibration of laminar airflow in Microbiology Laboratory.

3.0              RESPONSIBILITY

3.1              Microbiologist and above shall be responsible for the preparation and execution of this SOP

3.2              QC & QA Head shall be responsible for the checking and approval of this SOP.

4.0              ACCOUNTABILITY

Head QC

5.0              PROCEDURE



5.1              Cleaning

5.1.1        Ensure that the power supply to the LAF is switched ‘OFF’ before cleaning. Plug shall be removed from the socket.

5.1.2        The working bench surface of LAF shall be mop with freshly prepared 70 % IPA with a clean non-fiber shedding cloth.

5.1.3        Outer and inside surface of LAF shall be mop with 70 % IPA solution with non-fiber shedding cloth.

5.1.4        Frequency: Before and after any work.

5.2              Cleaning of filters:

5.2.1        Laminar airflow has two filters i.e. Pre-filter and HEPA filter (High-Efficiency Particulate Air Filter).

5.2.2        In case the Magnehelic gauge indicates the filter chocking, the Pre-filter shall be cleaned using a vacuum cleaner.

5.2.3        Pre-filter shall be checked once in three months for accumulation of dust.

5.2.4        Pre-filters shall be washed with a mild detergent and dried, if excessively dirty.

5.2.5        HEPA filter shall not be touched or opened by the laboratory personnel.

5.2.6        An authorized service engineer who shall also check for any rupture of the HEPA filters shall be  service the  LAF bench annually.

5.3              Operation

5.3.1        Ensure for the cleanliness of the instrument.

5.3.2        Ensure that Switch “ON” the power supply.

5.3.3        There are three switches for Air flow, U.V light, & normal light respectively and a magnehelic gauge on the panel of the instrument.

5.3.4        The knobs of the U.V. lamp and Airflow shall be turned to ‘ON’ position for 30 min. before start of work.

5.3.5        After 30 min. switch “OFF” the U.V lamp and work shall be started.

5.3.6        Ensure for the Magnehelic gauge, the level of a red oil  indicator should be between the 10.0 to 15.0 mm mark when the unidirectional flow is "ON".

5.3.7        After completion of work in LAF bench, again clean it with 70% IPA.

5.3.8        Ensure for the knobs of Airflow and light to 'OFF' position and put 'OFF' the main switch of LAF.

5.3.9        The details like U.V light burning hours, pressure, cleaning, instrument switch on/off shall be record.

5.4              Monitoring of microbial contamination

5.4.1        Media required Nutrient agar (NA)/Soyabean Casein Digest Agar (SCDA) for bacteria and Sabouraud's dextrose agar (SDA) for fungi.

5.4.2        Rehydrate the dehydrated media as per the manufacturers the instruction given on the bottle label and sterilize at 15 Lbs, 121 0C for 20 minutes or as per validated cycle Dispense about 20-25 ml of the media into sterile Petri dishes and allows to solidify.

5.4.3        Ensure that switch on the LAF & follow point 5.1.

5.4.4        Before working in LAF hands shall be rinsed with 70% IPAv/v and wear a nose mask and hand gloves.

5.4.5        Place 5 pre-incubated Nutrient Agar/SCDA plates and 5 pre-incubated Sabouraud's dextrose Agar plates, one on each corner and one in the center of the LAF bench.

                                                            L1                               L2

                                                                           L5

                                                            L4                               L3

 

5.4.6        Ensure that the plates shall be open with taking care not to introduce any extraneous contamination into the media and expose for 30 minutes.

5.4.7        After exposure, the plates shall be cover with the lids and number them appropriately so as to indicate the position of the plate on the LAF bench.

5.4.8        Media plates shall be incubated at 300C - 350C for 48-72 hours and further for 48-72 hrs at 20-25°C.

5.4.9        Also, incubate one plate of each media at the same temperature for the same time to serve as a negative control.

5.4.10    After incubation, ensure for the plates for growth and observation shall be recorded.

(as per Annexure - I).

5.4.11    If no bacterial or fungal growth is observed, then the LAF is OK.

5.4.12    If any growth is observed, repeat the monitoring twice.

5.4.13     If growth is still persistent, stop using the LAF with immediate effect and inform the Engineering  

department/supplier for rectification.

5.5              Frequency of monitoring

5.5.1         Once in 15 days and after every maintenance job by exposing a sterile media plate.

5.5.2         Calibration of air velocity & integrity testing of HEPA filter is done on contract every year.

6.0              REFERENCES

  N.A

7.0              ANNEXURE

Annexure No.

Title of the Annexure

Format No.

Annexure I

 Calibration record of LAF

 

8.0              ABBREVIATIONS :

SOP                                         :           Standard Operating Procedure

QCD                                       :           Quality Control Department

EQP                                        :           Equipment

Dept.                                       :           Department

QC                                          :           Quality Control

QA                                          :           Quality Assurance

QCD                                       :           Quality Control Department

No.                                          :           Number

N.A                                         :           Not Applicable

LAF                                        :           Laminar Air Flow

I.P.A                                       :           Isopropyl Alcohol

HEPA                                     :           High Efficient Particulate Air

SCDA                                     :           Soybean Casein Digest Agar


ANNEXURE: CALIBRATION RECORD OF LAMINAR AIR FLOW BY PLATE COUNT METHOD


Inst.ID:-

Make:-

Date

Location

Result after48 hrs at 30 - 350 C (cfu)

Result after further 72 hrs at 20 - 250 C (cfu)

Done By

 

 

 

 

 

 

 

 

 

 

 


Medium Used                       :

Plate Exposed For                : 30 Minutes

Plate Incubated At               : 32.5 °C  / 22.5 °C      +  2.5 °C

Plates Incubated For              : 48 Hrs. / 5 Days

Frequency                             : Once In 15 Days

 

Note: Limits :  No Growth Should Be Observed In Any Plate.

 

  Remark: The above mentioned laminar flow is working / not working satisfactorily.

SHARE

Owner

Hi. I’m Writer of Researchsop.com. ’ ’ Please share these SOPs to all concern pharma people for their development. I like to fullfill the need of curious people. These things inspire me to make things looks better.

  • Image
  • Image
  • Image
  • Image
  • Image
    Blogger Comment
    Facebook Comment

1 comments:

  1. Thank you for sharing this useful information. Laminar Airflow Manufacturers in Chennai are designing and producing equipment that ensures the controlled flow of air in environments such as laboratories, cleanrooms, hospitals, and manufacturing facilities. Laminar airflow systems are essential for maintaining sterile conditions by minimizing airborne contaminants, such as dust, microbes, and particles.

    ReplyDelete