SOP FOR OPERATION AND MAINTENANCE OF CAGE CHANGE STATION

1.0                      PURPOSE

To design a Standard Operating Procedure that describes the use and maintenance of the cage changing station that offers the benefits of vertical HEPA filtered airflow for product protection on the work surface.

2.0                      SCOPE

This Standard Operating Procedure (SOP) is applicable for the operation, and maintenance of the cage changing station for cage wash and animal care personnel at the animal house facility (AHF).

3.0                      RESPONSIBILITY

3.1         It is the responsibility of the facility manager to ensure that all portable equipment is appropriately cleaned, and maintained in good working order.

3.2                    It is the responsibility of the facility manager to ensure that all animal care staff is adequately trained and experienced in the use of the cage changing station.

3.3           It is the responsibility of the animal care staff using a cage changing station to read and understand the instruction manual and this SOP prior to equipment use.


                      




4.0                      DISTRIBUTION

4.1                The Quality Assurance department is responsible to keep SOP ‘Master Copy’ approved through the Quality Manager.

4.2                      The copy of ‘Control Copy’ of SOP is being distributed in the AHF department and placed near related Equipment/ Instrument as Display Copy. 

5.0                      DEFINITION(S) 

5.1                      Definition(s)

5.1.1       Cage changing station offers the benefits of vertical HEPA filtered airflow for product protection on the work surface.

6.0             ABBREVIATION(S)

6.1                      Abbreviations

6.1.1                 SOP           : Standard Operating Procedure

6.1.2                 AHF           : Animal House Facility

6.1.3                 AHI            : Animal House Instrument

6.1.4                 PPEs          : Personal Protective Equipment

6.1.5                 HEPA        :High Efficiency Particulate Air

7.0                      PROCEDURE

7.1.1          The blower and lights shall be turned and allow the blower to operate for a minimum of 10 minutes before manipulations are begun in the cabinet.

7.1.2                 Essential items/supplies shall only be allowed in the workstation.

7.1.3            Cabinet’s interior surface shall be decontaminated by wiping with a chemical disinfectant prior to commencing work. Chlorinated or halogen-based cleaners shall not be used unless followed by a final wipe/rinse with 70% alcohol.

7.1.4                 To reduce airflow disruptions:

7.1.5                 Movements in and out of the work area shall be minimal.

7.1.6                 Activity in the room, including opening and closing doors shall be minimal.

7.1.7              The following completion of all work, the station shall allow for 2-3 minute without activity to purge the unit.

7.1.8                 The cleaning/decontamination of the interior surfaces shall be repeated after the removal of all work materials by wiping with a chemical disinfectant.

7.1.9                 The blowers and lights shall be turned off. The cabinet shall not be used as a depository for excess laboratory equipment during periods of non-operation.

7.2                      Maintenance

7.2.1              Inspect the condition of the unit and electrical cord/plug to ensure safe operation. Equipment determined to be unsafe shall be removed from service immediately.

7.2.2        The station’s interior consists of both powder coat urethane and type 304 stainless steel surfaces. Chlorinated or halogen-based cleaners shall not be used unless followed by a final wipe/rinse with 70% alcohol.

7.2.3         The exterior surface shall be easily clean with alcohol followed by any mild household detergent.

7.2.4                 The exhaust pre-filter mesh netting shall be checked daily and cleaned as often as necessary, but at least weekly.

7.2.5                 The supply and exhaust pre-filter(s) shall be checked weekly and cleaned/changed as often as needed.

7.2.6        HEPA filters shall be replaced by qualified personnel when airflow velocity cannot be maintained.

7.2.7           Castors shall be checked yearly (older units have castors that screw onto the base these may loosen over time and eventually snap off).

7.2.8                 Certification shall be performed on a triennial basis by qualified personnel and shall include:

7.2.9                 Verifying pre-filter and HEPA filter integrity.

7.2.9.1            Measuring down flow velocity.

7.2.9.2            Checking exhaust motor/blower function.

7.2.10              Cleaning, sanitization & filter replacement

7.2.10.1         Unit work surface and pre-filters shall be thoroughly cleaned and sanitized at least weekly.

7.2.10.2         With the fan running, the work surface shall be cleaned with Oxivir Tb or Sporicidin wipes while wearing gloves, lab coats, and sleeves.

7.2.10.3         Lift the work surface and wipe the sides, underside, top & bottom of the stainless steel pre-filter cover, and bottom area of changing station. Use as many Oxivir Tb or Sporicidin wipes as necessary. Work surfaces that are extremely dirty shall be removed to cage wash for sanitation.

7.2.10.4         Units may contain two black foam filters, stacked, 1 coarse and 1 fine, or a single disposable white fiber filter.

7.2.10.5        Exhaust pre-filters shall be inspected weekly and cleaned/replaced when needed as described below

7.2.10.5.1    Turn Fan OFF.

7.2.10.5.2    Remove everything from the area that could possibly fall into the fan blades/motor.

7.2.10.5.3    Spritz the surface of the filter with Oxivir Tb or Sporicidin lightly.

7.2.10.5.4    Remove the filter from the housing.

7.2.10.5.5    Spritz underside of the filter with Oxivir Tb or Sporicidin.

7.2.10.5.6    Place filter in plastic bag and seal.

7.2.10.5.7    Foam filters are taken to cage wash for sanitization.

7.2.10.5.8    Fiber filters are disposable and are discarded.

7.2.10.6         Replace clean/new pre-filter(s) in unit.

7.2.10.7         Replace cleaned work surface.

7.2.10.8         Weekly cleaning of the unit and pre-filter inspection are noted on the Room Status Sheet.

8.0                      PRECAUTIONS

8.1                    All personal protective equipment like apron, gloves, mask, head caps, and shoe cover should be worn before using the instrument.

8.2                      Clean the equipment with a dry piece of robe or a lightly moistened one, Do not use organic solutions,

8.3                      Never use abrasive products or dissolvents. 

8.4                      Never pour water or liquids on the equipment.

8.5                      Once you have finished using the equipment turn it off by using the main switch, clean, and check the equipment so that it is in optimal conditions when you use it again.

9.0                      REFERENCES 

9.1                      Reference

9.1.1                 Refer to the manufacturer’s manual for additional information


END OF THE DOCUMENT

You may like to read these links: 

1. List of In-vivo laboratory Vendors (Capex)

2. List of SOPs and Documents for CPCSEA

3. List of All SOPs and Documents for Laboratory Instruments and Equipment

4. List of All SOPs and Documents for Animal House Facility

5. List of All SOPs and Documents for In-vivo Laboratory



SHARE

Owner

Hi. I’m Writer of Researchsop.com. ’ ’ Please share these SOPs to all concern pharma people for their development. I like to fullfill the need of curious people. These things inspire me to make things looks better.

  • Image
  • Image
  • Image
  • Image
  • Image
    Blogger Comment
    Facebook Comment

0 comments:

Post a Comment